Fa   |   Ar   |   En
   DEVELOPMENT AND VALIDATION OF HIGH PERFORMANCE LIQUID CHROMATOGRAPHY METHOD FOR ANALYSIS OF SIBUTRAMINE HYDROCHLORIDE AND ITS IMPURITY  
   
نویسنده CHANDORKAR JG ,KOTWAL VB ,DHANDE NS ,PACHPOR MP ,PANDE VV
منبع pakistan journal of pharmaceutical sciences - 2008 - دوره : 21 - شماره : 2 - صفحه:121 -124
چکیده    A simple, precise, rapid reproducible and selective reverse phase hplc method has been developed for the estimation of sibutramine hydrochloride monohydrate and its impurity in bulk as well as formulation. the analyte was resolved by using mobile phase (sodium dihydrogen phosphate and acetonitrile) at the flow rate of 1.0 ml/min. on isocratic hplc system consisting of jasco make uv visible detector of model uv 1575 & jasco make hplc pump of model pu 1580.an ods c- 8 rp column (4.6mm id, 250mm l, particle size 5 micron, at wavelength of 230 nm.
کلیدواژه Sibutramine hydrochloride ,impurity*- 5-(2R)-2-aminopropyl)-2-methoxybenzene sulphonamide
آدرس Aarti Drugs Ltd., India, Aarti Drugs Ltd., India, Aarti Drugs Ltd., India, Aarti Drugs Ltd., India, Aarti Drugs Ltd., India
 
     
   
Authors
  
 
 

Copyright 2023