Corrosion behavior of coated and uncoated bio implants in SBF (simulated body fluid)
|
|
|
|
|
نویسنده
|
iqbal w. ,zahra n. ,alam s. ,habib f. ,irfan m.
|
منبع
|
journal of the chemical society of pakistan - 2013 - دوره : 35 - شماره : 3 - صفحه:663 -665
|
چکیده
|
Surgical implants used in medical applications are basically the specific type of stainless steel materials. stainless steel has been used widely and successfully for various types of trauma and orthopedic reconstructions. if an uncoated (bare) stainless steel metal piece is implanted in any part of the body,it will get corrode in simulated body fluid (sbf) present inside the human body (a mixture of different salts). to overcome this problem a coating of titanium nitride (tin) was developed on stainless steel bio-implants using physical vapor deposition (pvd) method. both coated and uncoated implants were kept dipped in simulated body fluid for five months. the samples were removed and tested for corrosion life assessment after every fifteen days using weight loss method.
|
کلیدواژه
|
Chromium oxide; Corrosion; Simulated Body Fluid; Titanium Nitride
|
آدرس
|
pakistan institute of technology for minerals and advanced engineering materials,pcsir laboratories complex,ferozpur road, Pakistan, pakistan institute of technology for minerals and advanced engineering materials,pcsir laboratories complex,ferozpur road, Pakistan, pakistan institute of technology for minerals and advanced engineering materials,pcsir laboratories complex,ferozpur road, Pakistan, pakistan institute of technology for minerals and advanced engineering materials,pcsir laboratories complex,ferozpur road, Pakistan, pakistan institute of technology for minerals and advanced engineering materials,pcsir laboratories complex,ferozpur road, Pakistan
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Authors
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|