Dosimetry parameters of the IRH10 192Ir high dose rate brachytherapy source
|
|
|
|
|
نویسنده
|
yi c.-y. ,chun k.j. ,hah s.-h. ,kim h.-m. ,lim c.
|
منبع
|
journal of radiation research - 2010 - دوره : 51 - شماره : 4 - صفحه:485 -492
|
چکیده
|
The dosimetry parameters of the irh10 192ir high dose rate brachytherapy source were obtained from the dose calculation formalism recommended in the aapm task group no. 43 report using the monte carlo code penelope. the absorbed doses to water and air originating from the photons of the irh10 192ir brachytherapy source were calculated by the collision kerma approximation. the dose rate constant was evaluated to be (1.110 ± 0.011) cgy/h u -1ir. the dose rate per unit air kerma strength around the 192ir irh10 brachytherapy source and the anisotropy function were given in tables and figures.
|
|
|
آدرس
|
center for ionizing radiation,korea research institute of standards and science,doryong-dong,yuseong-gu, South Korea, center for ionizing radiation,korea research institute of standards and science,doryong-dong,yuseong-gu, South Korea, center for ionizing radiation,korea research institute of standards and science,doryong-dong,yuseong-gu, South Korea, center for ionizing radiation,korea research institute of standards and science,doryong-dong,yuseong-gu, South Korea, radiation safety division,national institute of food and drug safety evaluation,194 tongilro,eunpyeong-gu, South Korea
|
|
|
|
|
|
|
|
|
|
|
|
|
|