   A comparison of the dose distributions between the brachytherapy 125I source models,STM1251 and Oncoseed 6711,in a geometry lacking radiation equilibrium scatter conditions  
   
نویسنده tanaka k. ,kamo k.-i. ,tateoka k. ,asanuma o. ,sato k. ,takeda h. ,sakata k.-i. ,takada j.
منبع journal of radiation research - 2015 - دوره : 56 - شماره : 2 - صفحه:366 -371
چکیده    The purpose of this study was to estimate the uncertainty in the dose distribution for the 125i source stm1251,as measured with a radiophotoluminescent glass rod dosimeter and calculated using the monte carlo code egs5 in geometry that included the source structure reported by kirov et al. this was performed at a range of positions in and on a water phantom 18 cm in diameter and 16 cm in length. some dosimetry positions were so close to the surface that the backscatter margin was insufficient for photons. consequently,the combined standard uncertainty (csu) at the coverage factor k of 1 was 11.0-11.2% for the measurement and 1.8-3.6% for the calculation. the calculation successfully reproduced the measured dose distribution within 13%,with csu at k ≤ 1.6 (p > 0.3). dose distributions were then compared with those for the 125i source oncoseed 6711. our results supported the american association of physicists in medicine task group no. 43 updated protocol (tg43u1) formalism,in which stm1251 dose distributions were more penetrating than those of oncoseed 6711. this trend was also observed in the region near the phantom surface lacking the equilibrium radiation scatter conditions. in this region,the difference between the tg43u1 formalism and the measurement and calculation performed in the present study was not significant (p > 0.3) for either of the source models. selection of the source model based on the treatment plans according to the tg43u1 formalism will be practical. © the author 2015.
کلیدواژه 125I; Brachytherapy; EGS5; Glass rod dosimeter; TG43U1
آدرس division of physics,department of liberal arts and sciences,center of medical education,sapporo medical university,south 1,west 17,chuo-ward,sapporo,hokkaido,060-8556,japan,quantum energy applications,graduate school of engineering,hiroshima university,1-4-1 kagamiyama,higashi-hiroshima,hiroshima,739-8527, Japan, division of physics,department of liberal arts and sciences,center of medical education,sapporo medical university,south 1,west 17,chuo-ward,sapporo,hokkaido,060-8556, Japan, division of physics,department of liberal arts and sciences,center of medical education,sapporo medical university,south 1,west 17,chuo-ward,sapporo,hokkaido,060-8556, Japan, sapporo medical university hospital,south 1,west 15,chuo-ward,sapporo,hokkaido,060-8556, Japan, sapporo medical university hospital,south 1,west 15,chuo-ward,sapporo,hokkaido,060-8556, Japan, sapporo medical university hospital,south 1,west 15,chuo-ward,sapporo,hokkaido,060-8556, Japan, division of physics,department of liberal arts and sciences,center of medical education,sapporo medical university,south 1,west 17,chuo-ward,sapporo,hokkaido,060-8556, Japan, division of physics,department of liberal arts and sciences,center of medical education,sapporo medical university,south 1,west 17,chuo-ward,sapporo,hokkaido,060-8556, Japan
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved