comparison of maximal rubidium-82 activities for myocardial blood flow quantification between digital and conventional pet systems
|
|
|
|
|
نویسنده
|
dijk joris d. van ,jager pieter l. ,osch jochen a. c. van ,khodaverdi maryam ,dalen jorn a. van
|
منبع
|
journal of nuclear cardiology - 2019 - دوره : 26 - شماره : 4 - صفحه:1286 -1291
|
چکیده
|
Pet-based myocardial blood flow (mbf) quantification can be inaccurate when using high tracer activities. our aim was to derive the maximal rubidium-82 activity for mbf assessment using a new digital pet system and compare the results with conventional analog systems. 1.8 gbq rubidium-82 was injected into the cardiac insert of an anthropomorphic torso phantom. data were acquired for 10 min using an ingenuity tf (philips healthcare), discovery 690 (d690, ge healthcare), and digital pet prototype system (philips healthcare). the dynamic ranges, defined as the maximal measured activity in the reconstructed images deviating < 10% from the true present activity, were determined in all scans. the dynamic ranges were 312 mbq for ingenuity tf, 650 mbq for d690, and 654 mbq for digital pet prototype. the maximal rb-82 activity for mbf assessment using digital pet prototype is higher than that for its analog counterpart (ingenuity tf), but seems comparable to the d690.
|
کلیدواژه
|
digital pet ,direct photon counting ,dynamic range ,cardiac positron emission tomography ,rubidium-82
|
آدرس
|
isala hospital, department of nuclear medicine, the netherlands, isala hospital, department of nuclear medicine, the netherlands, isala hospital, department of medical physics, the netherlands, health systems, philips benelux, the netherlands, isala hospital, department of medical physics, the netherlands
|
|
|
|
|
|
|
|
|
|
|
|
|
|