>
Fa   |   Ar   |   En
   potential effects of low-dose average ct on cardiac implantable electronic devices  
   
نویسنده pan emily y. ,luo dershan ,pan tinsu
منبع journal of nuclear cardiology - 2019 - دوره : 26 - شماره : 4 - صفحه:1161 -1165
چکیده    Average ct has been shown to be more accurate than conventional helical ct in quantitation of the pet data. the risk of ct irradiation of a cardiac implantable electronic device (cied) causing an adverse event is low and is generally outweighed by the clinical benefit of a medically indicated examination. however, irradiation of cied over one breath cycle in cine ct scan for average ct could impose risks on a patient who is pacing dependent. the purpose of this study was to demonstrate that low-dose average ct can be safe for cied. a medtronic cied of model protecta vr was submerged in a saline bath for a series of 4-s cine ct scans on a ge ct scanner programmed to deliver a 2-cm-wide radiation at a dose rate of 0.9 to 41.2 mgy/s to the cied. the number of over-sensings was recorded as the interference of radiation to the cied. dose rates ≥ 1.9 mgy/s caused over-sensing. the higher the dose rate, the more over-sensings there were. the lowest dose rate of 0.9 mgy/s did not cause any over-sensing. low-dose average ct at 0.9 mgy/s can be safe for a cied patient who is pacing dependent.
کلیدواژه pet/ct imaging ,computed tomography (ct) ,myocardial perfusion imaging: pet
آدرس the university of texas md anderson cancer center, department of imaging physics, usa, the university of texas md anderson cancer center, department of radiation physics, usa, the university of texas md anderson cancer center, department of imaging physics, usa
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved