   Monitoring the respiratory rate by miniature motion sensors in premature infants: a comparative study  
   
نویسنده Kohn S ,Waisman D ,Pesin J ,Faingersh A ,Klotzman I C ,Levy C ,Hirshberg G ,Rotschild A ,Landesberg A
منبع journal of perinatology - 2016 - دوره : 36 - شماره : 2 - صفحه:116 -120
چکیده    Objective:existing respiratory rate (rr) monitors suffer from inaccuracy. the study assesses the accuracy of a novel modality that monitors lung ventilation with miniature motion sensors.study design:rr was measured by three methods: impedance technology, motion sensors and visual count, in babies (n=9) that breathed spontaneously or with respiratory support and babies (n=12) that received high-frequency oscillatory ventilation (hfov).results:a line close to equality (slope=0.96, r2=0.83) was obtained between the motion sensor and the visual count of the rr with narrow 95% limits of agreements (<14.0 b.p.m.). the relationship between the impedance and the visual count showed a lower correlation (r2=0.65) and wider 95% limits of agreements (21.4 b.p.m.). the motion sensor- and the ventilator-determined rrs demonstrated a good agreement during hfov, whereas the impedance failed to measure the rr during hfov.conclusion:monitoring rr with motion sensors is more accurate compared with the impedance, in infants, in all ventilation modes.
آدرس Pneumedicare, Israel, Carmel Medical Center, Department of Neonatology, Israel. Faculty of Medicine, Israel, Faculty of Biomedical Engineering, Israel, Faculty of Biomedical Engineering, Israel, Mayer Children's Hospital, Department of Neonatology, Israel, Pneumedicare, Israel, Faculty of Biomedical Engineering, Israel, Carmel Medical Center, Department of Neonatology, Israel. Faculty of Medicine, Israel, Faculty of Biomedical Engineering, Israel
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved