   MRI active guidewire with an embedded temperature probe and providing a distinct tip signal to enhance clinical safety  
   
نویسنده sonmez m. ,saikus c.e. ,bell j.a. ,franson d.n. ,halabi m. ,faranesh a.z. ,ozturk c. ,lederman r.j. ,kocaturk o.
منبع journal of cardiovascular magnetic resonance - 2012 - دوره : 14 - شماره : 1
چکیده    Background: the field of interventional cardiovascular mri is hampered by the unavailability of active guidewires that are both safe and conspicuous. heating of conductive guidewires is difficult to predict in vivo and disruptive to measure using external probes. we describe a clinical-grade 0.035 (0.89 mm) guidewire for mri right and left heart catheterization at 1.5 t that has an internal probe to monitor temperature in real-time,and that has both tip and shaft visibility as well as suitable flexibility. methods: the design has an internal fiberoptic temperature probe,as well as a distal solenoid to enhance tip visibility on a loopless antenna. we tested different tip-solenoid configurations to balance heating and signal profiles. we tested mechanical performance in vitro and in vivo in comparison with a popular clinical nitinol guidewire. results: the solenoid displaced the point of maximal heating (hot spot) from the tip to a more proximal location where it can be measured without impairing guidewire flexion. probe pullback allowed creation of lengthwise guidewire temperature maps that allowed rapid evaluation of design prototypes. distal-only solenoid attachment offered the best compromise between tip visibility and heating among design candidates. when fixed at the hot spot,the internal probe consistently reflected the maximum temperature compared external probes. real-time temperature monitoring was performed during porcine left heart catheterization. heating was negligible using normal operating parameters (flip angle,45°; sar,1.01 w/kg); the temperature increased by 4.2°c only during high rf power mode (flip angle,90°; sar,3.96 w/kg) and only when the guidewire was isolated from blood cooling effects by an introducer sheath. the tip flexibility and in vivo performance of the final guidewire design were similar to a popular commercial guidewire. conclusions: we integrated a fiberoptic temperature probe inside a 0.035 mri guidewire. real-time monitoring helps detect deleterious heating during use,without impairing mechanical guidewire operation,and without impairing mri visibility. we therefore need not rely on prediction to ensure safe clinical operation. future implementations may modulate specific absorption rate (sar) based on temperature feedback. © 2012 sonmez et al.; licensee biomed ltd.
آدرس division of intramural research,national heart lung and blood institute,national institutes of health,bethesda,md 20892,united states,institute of biomedical engineering,bogazici university,istanbul, Turkey, division of intramural research,national heart lung and blood institute,national institutes of health,bethesda,md 20892, United States, division of intramural research,national heart lung and blood institute,national institutes of health,bethesda,md 20892, United States, division of intramural research,national heart lung and blood institute,national institutes of health,bethesda,md 20892, United States, division of intramural research,national heart lung and blood institute,national institutes of health,bethesda,md 20892, United States, division of intramural research,national heart lung and blood institute,national institutes of health,bethesda,md 20892, United States, institute of biomedical engineering,bogazici university,istanbul, Turkey, division of intramural research,national heart lung and blood institute,national institutes of health,bethesda,md 20892, United States, division of intramural research,national heart lung and blood institute,national institutes of health,bethesda,md 20892, United States
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved