   Modified look-locker inversion recovery T1 mapping indices: Assessment of accuracy and reproducibility between magnetic resonance scanners  
   
نویسنده raman f.s. ,kawel-boehm n. ,gai n. ,freed m. ,han j. ,liu c.-y. ,lima j.a. ,bluemke d.a. ,liu s.
منبع journal of cardiovascular magnetic resonance - 2013 - دوره : 15 - شماره : 1
چکیده    Background: cardiovascular magnetic resonance (cmr) t1 mapping indices,such as t1 time and partition coefficient (λ),have shown potential to assess diffuse myocardial fibrosis. the purpose of this study was to investigate how scanner and field strength variation affect the accuracy and precision/reproducibility of t1 mapping indices. methods. cmr studies were performed on two 1.5t and three 3t scanners. eight phantoms were made to mimic the t1/t2 of pre- and post-contrast myocardium and blood at 1.5t and 3t. t1 mapping using molli was performed with simulated heart rate of 40-100 bpm. inversion recovery spin echo (ir-se) was the reference standard for t1 determination. accuracy was defined as the percent error between molli and ir-se,and scan/re-scan reproducibility was defined as the relative percent mean difference between repeat molli scans. partition coefficient was estimated by δr1myocardium phantom/δr1blood phantom. generalized linear mixed model was used to compare the accuracy and precision/reproducibility of t1 and λ across field strength,scanners,and protocols. results: field strength significantly affected molli t1 accuracy (6.3% error for 1.5t vs. 10.8% error for 3t,p<0.001) but not λ accuracy (8.8% error for 1.5t vs. 8.0% error for 3t,p=0.11). partition coefficients of molli were not different between two 1.5t scanners (47.2% vs. 47.9%,p=0.13),and showed only slight variation across three 3t scanners (49.2% vs. 49.8% vs. 49.9%,p=0.016). partition coefficient also had significantly lower percent error for precision (better scan/re-scan reproducibility) than measurement of individual t1 values (3.6% for λ vs. 4.3%-4.8% for t1 values,approximately,for pre/post blood and myocardium values). conclusion: based on phantom studies,t1 errors using molli ranged from 6-14% across various mr scanners while errors for partition coefficient were less (6-10%). compared with absolute t1 times,partition coefficient showed less variability across platforms and field strengths as well as higher precision. © 2013 raman et al.; licensee biomed central ltd.
کلیدواژه Diffuse myocardial fibrosis; Extracellular volume fraction (ECV); Modified look-locker with inversion recovery (MOLLI); Partition coefficient (λ); T1 mapping
آدرس radiology and imaging sciences,national institutes of health clinical center,10 center drive,msc-1182,bethesda,md 20892-1182, United States, radiology and imaging sciences,national institutes of health clinical center,10 center drive,msc-1182,bethesda,md 20892-1182,united states,clinic of radiology and nuclear medicine,university of basel hospital,basel, Switzerland, radiology and imaging sciences,national institutes of health clinical center,10 center drive,msc-1182,bethesda,md 20892-1182, United States, new york university school of medicine,new york,ny, United States, u.s. food and drug administration,rockville,md, United States, radiology and imaging sciences,national institutes of health clinical center,10 center drive,msc-1182,bethesda,md 20892-1182, United States, division of cardiology,johns hopkins university school of medicine,baltimore,md, United States, radiology and imaging sciences,national institutes of health clinical center,10 center drive,msc-1182,bethesda,md 20892-1182,united states,molecular biomedical imaging laboratory,national institute of biomedical imaging and bioengineering,bethesda,md, United States, radiology and imaging sciences,national institutes of health clinical center,10 center drive,msc-1182,bethesda,md 20892-1182,united states,molecular biomedical imaging laboratory,national institute of biomedical imaging and bioengineering,bethesda,md, United States
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved