   Impact of motion correction on reproducibility and spatial variability of quantitative myocardial T2 mapping  
   
نویسنده roujol s. ,basha t.a. ,weingärtner s. ,akçakaya m. ,berg s. ,manning w.j. ,nezafat r.
منبع journal of cardiovascular magnetic resonance - 2015 - دوره : 17 - شماره : 1
چکیده    Background: to evaluate and quantify the impact of a novel image-based motion correction technique in myocardial t2 mapping in terms of measurement reproducibility and spatial variability. methods: twelve healthy adult subjects were imaged using breath-hold (bh),free breathing (fb),and free breathing with respiratory navigator gating (fb∈+∈nav) myocardial t2 mapping sequences. fifty patients referred for clinical cmr were imaged using the fb∈+∈nav sequence. all sequences used a t2 prepared (t2prep) steady-state free precession acquisition. in-plane myocardial motion was corrected using an adaptive registration of varying contrast-weighted images for improved tissue characterization (arctic). dice similarity coefficient (dsc) and myocardial boundary errors (mbe) were measured to quantify the motion estimation accuracy in healthy subjects. t2 mapping reproducibility and spatial variability were evaluated in healthy subjects using 5 repetitions of the fb∈+∈nav sequence with either 4 or 20 t2prep echo times (te). subjective t2 map quality was assessed in patients by an experienced reader using a 4-point scale (1-non diagnostic,4-excellent). results: arctic led to increased dsc in bh data (0.85∈±∈0.08 vs. 0.90∈±∈0.02,p∈=∈0.007),fb data (0.78∈±∈0.13 vs. 0.90∈±∈0.21,p∈<∈0.001),and fb∈+∈nav data (0.86∈±∈0.05 vs. 0.90∈±∈0.02,p∈=∈0.002),and reduced mbe in bh data (0.90∈±∈0.40 vs. 0.64∈±∈0.19 mm,p∈=∈0.005),fb data (1.21∈±∈0.65 vs. 0.63∈±∈0.10 mm,p∈<∈0.001),and fb∈+∈nav data (0.81∈±∈0.21 vs. 0.63∈±∈0.08 mm,p∈<∈0.001). improved reproducibility (4te: 5.3∈±∈2.5 ms vs. 4.0∈±∈1.5 ms,p∈=∈0.016; 20te: 3.9∈±∈2.3 ms vs. 2.2∈±∈0.5 ms,p∈=∈0.002),reduced spatial variability (4te: 12.8∈±∈3.5 ms vs. 10.3∈±∈2.5 ms,p∈<∈0.001; 20te: 9.7∈±∈3.5 ms vs. 7.5∈±∈1.4 ms) and improved subjective score of t2 map quality (3.43∈±∈0.79 vs. 3.69∈±∈0.55,p∈<∈0.001) were obtained using arctic. conclusions: the arctic technique substantially reduces spatial mis-alignment among t2-weighted images and improves the reproducibility and spatial variability of in-vivo t2 mapping. © 2015 roujol et al.
کلیدواژه Image registration; Motion correction; Myocardial T2 mapping; Quantitative myocardial tissue characterization
آدرس department of medicine (cardiovascular division),beth israel deaconess medical center,330 brookline ave,boston,ma 02215, United States, department of medicine (cardiovascular division),beth israel deaconess medical center,330 brookline ave,boston,ma 02215,united states,biomedical engineering department,cairo university,giza, Egypt, department of medicine (cardiovascular division),beth israel deaconess medical center,330 brookline ave,boston,ma 02215,united states,computer assisted clinical medicine,university medical center mannheim,heidelberg university,mannheim, Germany, department of medicine (cardiovascular division),beth israel deaconess medical center,330 brookline ave,boston,ma 02215, United States, department of medicine (cardiovascular division),beth israel deaconess medical center,330 brookline ave,boston,ma 02215, United States, department of medicine (cardiovascular division),beth israel deaconess medical center,330 brookline ave,boston,ma 02215,united states,radiology,beth israel deaconess medical center,harvard medical school,boston,ma, United States, department of medicine (cardiovascular division),beth israel deaconess medical center,330 brookline ave,boston,ma 02215, United States
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved