Myocardial T1-mapping at 3T using saturation-recovery: reference values,precision and comparison with MOLLI
|
|
|
|
|
نویسنده
|
weingärtner s. ,meßner n.m. ,budjan j. ,loßnitzer d. ,mattler u. ,papavassiliu t. ,zöllner f.g. ,schad l.r.
|
منبع
|
journal of cardiovascular magnetic resonance - 2016 - دوره : 18 - شماره : 1 - صفحه:1 -9
|
چکیده
|
Background: myocardial t1-mapping recently emerged as a promising quantitative method for non-invasive tissue characterization in numerous cardiomyopathies. commonly performed with an inversion-recovery (ir) magnetization preparation at 1.5t,the application at 3t has gained due to increased quantification precision. alternatively,saturation-recovery (sr) t1-mapping has recently been introduced at 1.5t for improved accuracy. thus,the purpose of this study is to investigate the robustness and precision of sr t1-mapping at 3t and to establish accurate reference values for native t1-times and extracellular volume fraction (ecv) of healthy myocardium. methods: balanced steady-state free-precession (bssfp) saturation-pulse prepared heart-rate independent inversion-recovery (sapphire) and saturation-recovery single-shot acquisition (sasha) t1-mapping were compared with the modified look-locker inversion recovery (molli) sequence at 3t. accuracy and precision were studied in phantom. native and post-contrast t1-times and regional ecv were determined in 20 healthy subjects (10 men,27 ± 5 years). subjective image quality,susceptibility artifact rating,in-vivo precision and reproducibility were analyzed. results: sr t1-mapping showed <4 % deviation from the spin-echo reference in phantom in the range of t1 = 100-2300 ms. the average quality and artifact scores of the t1-mapping methods were: molli:3.4/3.6,sapphire:3.1/3.4,sasha:2.9/3.2; (1: poor - 4: excellent/1: strong - 4: none). sapphire and sasha yielded significantly higher t1-times (sapphire: 1578 ± 42 ms,sasha: 1523 ± 46 ms),in-vivo t1-time variation (sapphire: 60.1 ± 8.7 ms,sasha: 70.0 ± 9.3 ms) and lower ecv-values (sapphire: 0.20 ± 0.02,sasha: 0.21 ± 0.03) compared with molli (t1: 1181 ± 47 ms,ecv: 0.26 ± 0.03,precision: 53.7 ± 8.1 ms). no significant difference was found in the inter-subject variability of t1-times or ecv-values (t1: p = 0.90,ecv: p = 0.78),the observer agreement (inter: p > 0.19; intra: p > 0.09) or consistency (inter: p > 0.07; intra: p > 0.17) between the three methods. conclusions: saturation-recovery t1-mapping at 3t yields higher accuracy,comparable inter-subject,inter- and intra-observer variability and less than 30 % precision-loss compared to molli. © 2016 the author(s).
|
کلیدواژه
|
3T; Cardiovascular magnetic resonance; MOLLI; Reference values; SAPPHIRE; SASHA; Saturation-recovery T1-mapping
|
آدرس
|
computer assisted clinical medicine,university medical center mannheim,medical faculty mannheim,heidelberg university,theodor-kutzer-ufer 1-3,mannheim,68167,germany,electrical and computer engineering,university of minnesota,minneapolis,mn,united states,center for magnetic resonance research,university of minnesota,minneapolis,mn, United States, computer assisted clinical medicine,university medical center mannheim,medical faculty mannheim,heidelberg university,theodor-kutzer-ufer 1-3,mannheim,68167,germany,dzhk (german centre for cardiovascular research) partner site heidelberg/mannheim,mannheim, Germany, institute of clinical radiology and nuclear medicine,university medical center mannheim,medical faculty mannheim,heidelberg university,mannheim, Germany, 1st department of medicine cardiology,university medical center mannheim,medical faculty mannheim,heidelberg university,mannheim, Germany, institute of clinical radiology and nuclear medicine,university medical center mannheim,medical faculty mannheim,heidelberg university,mannheim, Germany, dzhk (german centre for cardiovascular research) partner site heidelberg/mannheim,mannheim,germany,1st department of medicine cardiology,university medical center mannheim,medical faculty mannheim,heidelberg university,mannheim, Germany, computer assisted clinical medicine,university medical center mannheim,medical faculty mannheim,heidelberg university,theodor-kutzer-ufer 1-3,mannheim,68167, Germany, computer assisted clinical medicine,university medical center mannheim,medical faculty mannheim,heidelberg university,theodor-kutzer-ufer 1-3,mannheim,68167, Germany
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Authors
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|