   Accuracy of five electronic foramen locators with different operating systems: An ex vivo study  
   
نویسنده de vasconcelos b.c. ,bueno m.m. ,luna-cruz s.m. ,duarte m.a.h. ,fernandes c.a.o.
منبع journal of applied oral science - 2013 - دوره : 21 - شماره : 2 - صفحه:132 -137
چکیده    Objective: the aim of this study was to evaluate,ex vivo,the precision of five electronic root canal length measurement devices (erclmds) with different operating systems: the root zx,mini apex locator,propex ii,ipex,and romiapex a-15,and the possible influence of the positioning of the instrument tips short of the apical foramen. material and methods: forty-two mandibular bicuspids had their real canal lengths (rl) previously determined. electronic measurements were performed 1.0 mm short of the apical foramen (-1.0),followed by measurements at the apical foramen (0.0). the data resulting from the comparison of the erclmd measurements and the rl were evaluated by the wilcoxon and friedman tests at a significance level of 5%. results: considering the measurements performed at 0.0 and -1.0,the precision rates for the erclmds were: 73.5% and 47.1% (root zx),73.5% and 55.9% (mini apex locator),67.6% and 41.1% (propex ii),61.7% and 44.1% (ipex),and 79.4% and 44.1% (romiapex a-15),respectively,considering ±0.5 mm of tolerance. regarding the mean discrepancies,no differences were observed at 0.0; however,in the measurements at -1.0,the ipex,a multi-frequency erclmd,had significantly more discrepant readings short of the apical foramen than the other devices,except for the propex ii,which had intermediate results. when the erclmds measurements at -1.0 were compared with those at 0.0,the propex ii,ipex and romiapex a-15 presented significantly higher discrepancies in their readings. conclusions: under the conditions of the present study,all the erclmds provided acceptable measurements at the 0.0 position. however,at the -1.0 position,the erclmds had a lower precision,with statistically significant differences for the propex ii,ipex,and romiapex a-15.
کلیدواژه Anatomy and Histology; Dental equipment; Electric impedance; Endodontics; Odontometry methods; Tooth apex
آدرس sobral school of dentistry,federal university of ceara and school of pharmacy,dentistry and nursing,federal university of ceara,fortaleza,ce, Brazil, brazilian dental association,ceará section,fortaleza,ce, Brazil, sobral school of dentistry,federal university of ceará,fortaleza,ce, Brazil, department of operative dentistry,endodontics and dental materials,bauru school of dentistry,university of são paulo,bauru,sp, Brazil, school of pharmacy,dentistry and nursing,federal university of ceará,fortaleza,ce, Brazil
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved