   Accuracy and repeatability of refractive error measurements by photorefractometry  
   
نویسنده rajavi z. ,sabbaghi h. ,baghini a. ,yaseri m. ,sheibani k. ,norouzi g.
منبع journal of ophthalmic and vision research - 2015 - دوره : 10 - شماره : 3 - صفحه:221 -228
چکیده    Purpose: to determine the accuracy of photorefraction and autorefraction as compared to cycloautorefraction and to detect the repeatability of photorefraction. methods: this diagnostic study included the right eyes of 86 children aged 7-12 years. refractive status was measured using photorefraction (plusoptix so4,gmbh,nürnberg,germany) and autorefraction (topcon rm800,usa) with and without cycloplegia. photorefraction for each eye was performed three times to assess repeatability. results: the overall agreement between photorefraction and cycloautorefraction was over 81% for all refractive errors. photorefractometry had acceptable sensitivity and specificity for myopia and astigmatism. there was no statistically significant difference considering myopia and astigmatism in all comparisons,while the difference was significant for hyperopia using both amblyogenic (p = 0.006) and nonamblyogenic criteria (p = 0.001). a myopic shift of 1.21 diopter (d) and 1.58 d occurred with photorefraction in nonamblyogenic and amblyogenic hyperopia,respectively. using revised cut-off points of + 1.12 d and + 2.6 d instead of + 2.00 d and + 3.50 d improved the sensitivity of photorefractometry to 84.62% and 69.23%,respectively. the repeatability of photorefraction for measurement of myopia,astigmatism and hyperopia was acceptable (intra-cluster correlation [icc]: 0.98,0.94 and 0.77,respectively). autorefraction results were significantly different from cycloautorefraction in hyperopia (p < 0.0001),but comparable in myopia and astigmatism. also,noncycloglegic autorefraction results were similar to photorefraction in this study. conclusion: although photorefraction was accurate for measurement of myopia and astigmatism,its sensitivity for hyperopia was low which could be improved by considering revised cut-off points. considering cut-off points,photorefraction can be used as a screening method. © 2015 journal of ophthalmic and vision research | published by wolters kluwer - medknow.
کلیدواژه Accuracy; Autorefraction; Cycloautorefraction; Photorefraction; Repeatability
آدرس ophthalmic epidemiology research center,shahid beheshti university of medical sciences,tehran,iran,department of ophthalmology,shahid beheshti university of medical sciences,tehran,iran,basir eye safety research center,basir eye clinic, ایران, basir eye safety research center,basir eye clinic,tehran,iran,ophthalmic research center,shahid beheshti university of medical sciences,no. 23,paidarfard st., ایران, basir eye safety research center,basir eye clinic, ایران, ophthalmic research center,shahid beheshti university of medical sciences,no. 23,paidarfard st.,tehran,iran,department of epidemiology and biostatistics,tehran university of medical sciences, ایران, basir eye safety research center,basir eye clinic, ایران, shahid beheshti university of medical sciences, ایران
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved