   utilization of gafchromic ebt3 films for postal dose audits of 6 mv x ray beam calibration of linear accelerator  
   
نویسنده mutahari said mahdi ,tajmalzai abasin
منبع afghanistan journal of basic medical sciences - 2025 - دوره : 2 - شماره : 1 - صفحه:16 -23
چکیده    Background: periodic dose quality audits of medical linear accelerators are essential to ensure proper calibration and accurate dose delivery. we aimed to assess the accuracy of 6 mv x-ray beam calibration through a postal dose audit using gafchromic ebt3 films. methods: the experiment was conducted at rsud pasar minggu hospital in jakarta, indonesia, in october 2023. a reference calibration curve for gafchromic ebt3 films was created by irradiating 3×3 cm² films at a 10 cm depth in a water phantom with a 10×10 cm² field and 100 cm source-to-surface distance, using doses 0-300 cgy with a 6 mv x-ray beam from a varian trilogy medical linear accelerator at rsud pasar minggu hospital in jakarta, indonesia. the films were scanned with a microtek tma1600-iii scanner, and optical density values were plotted against delivered doses. three unexposed films were sent to the hospital for irradiation to 2 gy in october 2023, and the returned films were analyzed by comparing their optical density values to the calibration curve. the measured dose was 1.92 gy, a -4.2% deviation from the prescribed 2 gy dose, within the ±5% tolerance. a parallel verification using optically stimulated luminescence dosimeters confirmed the gafchromic results. this study demonstrates the feasibility of using gafchromic ebt3 films for accurate, efficient calibration of linear accelerator x-ray beams in postal audits. results: the measured dose was 1.92 gy, representing a -4.2% deviation from the prescribed 2 gy dose, within the accepted ±5% tolerance. a parallel verification using optically stimulated luminescence dosimeters showed agreement with the gafchromic results. conclusion: the study confirms the feasibility of using gafchromic ebt3 films for accurate and efficient calibration of linear accelerator x-ray beams in postal audits.
کلیدواژه gafchromic ebt3 film ,linear accelerator ,dose quality audit ,dose calibration curve
آدرس kabul university of medical science, department of radiology, afghanistan, kabul university of medical science, department of radiology, afghanistan
پست الکترونیکی sm_mutahari@kums.edu.af
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved