   a method to modify the response of neutron dosimeters in the fields with energy spectra different from the calibration field  
   
نویسنده moslehi amir ,baradaran samaneh ,rezaeian peiman
منبع journal of nuclear research and applications - 2021 - دوره : 1 - شماره : 1 - صفحه:1 -8
چکیده    One problem in neutron dosimetry is that dosimeter responses per unit dose-equivalent in measurement and calibration fields are different. hence, the neutron dose suffers a large uncertainty. in this work, a correction method is introduced to modify the dosimeters response in the fields with known energy spectra. to examine it, responses of thermoluminescence dosimeters (tlds) to fast neutrons in 241am-be (calibration), 252cf and 239pu-be fields for four personal dose-equivalents (hp (10) values) of 5, 10, 15 and 20 msv are measured. the results obtained reveal that the maximum differences of the original responses measured in 252cf and 239pu-be fields from 241am-be field are 15% and 42%, respectively. after correction, above differences reduce to 4.7% and 10.8%. finally, it can be concluded that the method proposed here, improves the accuracy of dose measurement in the neutron fields with known energy spectra.
کلیدواژه fast neutron spectrum ,response ,calibration field ,correction factor ,tld
آدرس atomic energy organization of iran (aeoi), radiation applications research school, nuclear science and technology research institute, iran, atomic energy organization of iran (aeoi), reactor and nuclear safety research school, nuclear science and technology research institute, iran, atomic energy organization of iran (aeoi), radiation applications research school, nuclear science and technology research institute, iran
پست الکترونیکی prezaeian@aeoi.org.ir
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved