fabrication and validation of respiratory control belt using load cell sensor
|
|
|
|
|
نویسنده
|
tootoonchi dallal zahra ,gholami somayeh ,salehi barough mehdi
|
منبع
|
frontiers in biomedical technologies - 2020 - دوره : 7 - شماره : 4 - صفحه:281 -285
|
چکیده
|
Purpose: in this work, we have focused on fabrication and validation of a respiratory belt that produces respiratory signals from the patient’s abdominal and thorax.materials and methods: a load cell transducer was attached to the belt to create an electrical signal from respiratory movement. it converts a force or load into an equivalent electrical signal or digitized load value. the accuracy of the signals from our respiratory control belt was evaluated according to a comparison with the signals from commercial somno medical device.results: the pattern of the signals from our respiratory belt is in good agreement (4%/3mm) with the pattern of commercial somno medical device.conclusion: the manufactured respiratory control belt can be used during imaging and radiotherapy procedures for breath control in a patient's abdominal and thorax region
|
کلیدواژه
|
respiratory getting belt ,load cell sensor ,breath control
|
آدرس
|
islamic azad university, central tehran branch, department of medical radiation engineering, iran, tehran university of medical sciences, cancer institute, school of medicine, department of radiation oncology, iran, islamic azad university, central tehran branch, department of medical radiation engineering, iran
|
پست الکترونیکی
|
salehi@gmail.com
|
|
|
|
|
|
|
|
|
|
|
|