>
Fa   |   Ar   |   En
   determination of diagnostic reference level in routine examinations of digital radiography in nigeria  
   
نویسنده fatile oluwanifesii ,ajayi oladele samuel ,arogunjo a.m. ,ibikunle s. b.
منبع radiation physics and engineering - 2024 - دوره : 5 - شماره : 3 - صفحه:35 -39
چکیده    While ionizing radiation plays a pivotal role in the precise diagnosis and treatment, it concurrently engenders risks, including an elevated incidence of cancer. the research speaks to the discernible decline in quality assurance programmes and dose measurement endeavors within nigerian imaging facilities, with a substantial portion lacking established protocols for routine machine calibration and dose measurement. this study encompasses a large-scale survey involving 307 adult patients undergoing routine x-ray procedures in two hospitals in nigeria. thermo-luminescent dosimeters (tlds) were used for measuring entrance skin dose (esd) of consenting adult patients. the mean esd values ranged from 1.16 mgy to a maximum of 3.94 mgy. notably, these values were predominantly below the dose reference levels (drls) established by reputable bodies such as nrpb, cec, iaea, and uk for most examinations. the main purpose of this study was to determine the diagnostic reference level (drl) for routine digital radiography examinations in nigeria.
کلیدواژه diagnostic reference level (drl) ,thermo-luminescent dosimeters (tlds) ,entrance skin dose (esd) ,radiography
آدرس federal university of technology, department of physics, nigeria, federal university of technology, department of physics, nigeria, federal university of technology, department of physics, nigeria, federal university of technology, department of physics, nigeria
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved