determination of coagulation time of in-vivo cut bleeding treated by nonthermal atmospheric pressure plasma
|
|
|
|
|
|
|
|
نویسنده
|
shahbazi rad zahra ,abbasi davani fereydoun ,etaati gholamreza ,seifi samaneh
|
|
منبع
|
radiation physics and engineering - 2020 - دوره : 1 - شماره : 2 - صفحه:7 -11
|
|
چکیده
|
The aim of this research was determination of the required time for coagulation of in vivo cut bleeding treated by non-thermal atmospheric pressure plasma. to meet this, an atmospheric pressure plasma jet device was designed and constructed. helium was used as working gas. the electrical parameters and optical emission spectrum of helium plasma were measured. the averaged treatment time to coagulate the incision bleeding on the mouse liver was obtained 8.6 µs, and the average time of naturally incision bleeding coagulation was 10 min.
|
|
کلیدواژه
|
helium plasma jet ,non-thermal atmospheric ,pressure plasma ,in vivo blood coagulation
|
|
آدرس
|
shahid beheshti university, radiation application department, iran, shahid beheshti university, radiation application department, iran, amirkabir university, physics and energy engineering department, iran, shahrood university of technology, physics faculty, nuclear physics department, iran
|
|