>
Fa   |   Ar   |   En
   Diagnostic Performance of Two Point-of-Care Tests for Anti-HCV Detection  
   
نویسنده Rosa Lígia da ,Dantas-Corrêa Esther Buzaglo ,Narciso-Schiavon Janaína Luz ,Schiavon Leonardo de Lucca
منبع hepatitis monthly - 2013 - دوره : 13 - شماره : 9 - صفحه:1 -5
چکیده    Background: besides the great importance of the issue in terms of public health, there is a lack of studies evaluating the performance of several of the currently used point of care tests (pocts) for the detection of anti-hcv. objectives: to investigate the performance of two pocts for anti-hcv detection and to assess the impact of the reading time on diagnostic performance. patients and methods: a total of 307 subjects were divided into three groups (1- hcv infected; 2- other chronic liver diseases; and 3-controls). the pocts hcv rapid test bioeasy® and imuno-rapido hcv® were read at 3, 5, 10, 15, 20 and 30 minutes. the sensitivity and specificity of the pocts were calculated in relation to anti-hcv detection by chemiluminescence. results: valid results were obtained for all tests. when compared to the chemiluminescence, both tests showed sensitivity of 97.1% and specificity of 100%. no changes in the sensitivity or specificity of the tests were observed at different reading times and when patients with other chronic liver diseases were evaluated as a control group. conclusions: the pocts evaluated in this study showed high sensitivity and specificity, with no change in the performance after the thirdminute of reading.
کلیدواژه Point-of-Care Systems; Hepatitis C; Diagnosis
آدرس Federal University of Santa Catarina, Division of Gastroenterology, Brazil, Federal University of Santa Catarina, Division of Gastroenterology, Brazil, Federal University of Santa Catarina, Division of Gastroenterology, Brazil, Federal University of Santa Catarina, Division of Gastroenterology, Brazil
پست الکترونیکی leo-jf@uol.com.br
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved