Corneal thickness measurements with Scheimpflug and slit scanning imaging techniques in keratoconus
|
|
|
|
|
نویسنده
|
aghazadeh amiri m. ,hashemi h. ,ramin s. ,yekta a. ,taheri a. ,nabovati p. ,khabazkhoob m.
|
منبع
|
journal of current ophthalmology - 2017 - دوره : 29 - شماره : 1 - صفحه:23 -27
|
چکیده
|
Purpose to determine the repeatability of corneal thickness measurements with scheimpflug (pentacam) and slit scanning (orbscan) imaging techniques in different grades of keratoconus. methods this study was conducted as a cross-sectional research. imaging with orbscan and pentacam was performed on patients with different grades of keratoconus. with each device,3 measurements were taken at 10 min intervals. repeatability indices in different grades of keratoconus were calculated for each device. results seventy-four eyes of 42 keratoconus patients were enrolled. repeatability index (ri) of central corneal thickness (cct) measurements in keratoconus grade 1,2,and 3,were 12.8,9.9,and 24.2 with pentacam,and 23.6,26.3,and 59.3 with orbscan,respectively. for the thinnest point,these figures were 9.6,8.0,and 35.7 with pentacam and 19.5,16.6,and 26.8 with orbscan,respectively. the 95% limit of agreement (loa) between pentacam and orbscan in measuring cct and thinnest point in grade 1 were −25.5–47.7 mic and −33.3–32.8 mic,respectively. these results for grade 2 were −9.8–50.6 mic and −26.2–43.7 mic,respectively. in grade 3,95% loa were −20–64.6 mic and −31.4–60.5 mic,respectively. conclusions the results of this study showed that although repeated measurements of the cct with orbscan and pentacam are strongly correlated,repeatability values of cct measurements significantly decrease at more advanced grades of keratoconus. in all keratoconus grades,repeatability of cct measurements was better with pentacam than orbscan. these findings indicate that corneal thickness readings have less validity in patients with advanced keratoconus. © 2016 iranian society of ophthalmology
|
کلیدواژه
|
Agreement; Corneal thickness; Keratoconus; Orbscan; Pentacam; Repeatability
|
آدرس
|
department of optometry,faculty of rehabilitation science,shahid beheshti university of medical sciences,tehran, ایران, noor research center for ophthalmic epidemiology,noor eye hospital,tehran,iran,noor ophthalmology research center,noor eye hospital,tehran, ایران, department of optometry,faculty of rehabilitation science,shahid beheshti university of medical sciences,tehran, ایران, department of optometry,school of paramedical sciences,mashhad university of medical sciences,mashhad, ایران, noor research center for ophthalmic epidemiology,noor eye hospital,tehran, ایران, noor ophthalmology research center,noor eye hospital,tehran, ایران, noor research center for ophthalmic epidemiology,noor eye hospital,tehran,iran,department of medical surgical nursing,school of nursing and midwifery,shahid beheshti university of medical sciences,tehran, ایران
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Authors
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|