An Unnoticed Broken Sheathed Metallic Stylet in an Endotracheal Tube: A case report
|
|
|
|
|
نویسنده
|
Sharma Pradeep K ,Khan Rashid M ,Kaul Naresh
|
منبع
|
sultan qaboos university medical journal - 2010 - دوره : 10 - شماره : 1 - صفحه:126 -128
|
چکیده
|
We report a 58-years old patient, who underwent surgery for a fracture to the neck of the femur. tracheal intubation was performed with the aid of a stylet; however, 15 minutes later, it was brought to the notice of the attending anesthesiologist that a broken piece of stylet had been left inside the lumen of the endotracheal tube. prior to this, there was no evidence of a foreign body in the endotracheal tube or tracheobronchial tree. the broken piece of stylet was successfully retrieved with the help of a kocher’s forceps
|
کلیدواژه
|
Broken stylet ,Endotracheal tube obstruction ,Foreign body ,Case report ,Oman
|
آدرس
|
National Trauma Centre, Khoula Hospital, Department of Anesthesia & ICU, Oman, National Trauma Centre, Khoula Hospital, Department of Anesthesia & ICU, Oman, National Trauma Centre, Khoula Hospital, Department of Anesthesia & ICU, Oman
|