Fa   |   Ar   |   En
   Microwave System for the Dielectric Analysis and Moisture Evaluation  
   
نویسنده El Gholb Youssef ,O’Leary Paul
منبع mediterranean telecommunications journal - 2016 - دوره : 6 - شماره : 1 - صفحه:45 -48
چکیده    A detailed method for determining the moisture level and the dielectric properties of materials using microwave free-space transmission/reflection measurements system and the modular hephaistos-system (high electromagnetic power heating automated injected structures oven system) is presented. steps involved in the determination process are explained and their effect on the s-parameters is presented. a through-reflect-line/match (trl/m) free-space calibration method was used, which greatly reduces system errors owing to its simple and stationary nature.
کلیدواژه TRL/M Calibration;Non-Destructive Testing measurement;Vector Network Analyser;Dielectric Materials;Complex Permittivity;Modular HEPHAISTOS-System
آدرس University of Bologna, Department of Electronic, Telecommunication and Information Technology, Italy, Waterford Institute of Technology, Engineering Department, Ireland
 
     
   