>
Fa   |   Ar   |   En
   Rapid and reliable calibration of laser beam deflection system for microcantilever-based sensor setups  
   
نویسنده mishra r. ,grange w. ,hegner m.
منبع journal of sensors - 2012 - دوره : 2012 - شماره : 0
چکیده    Cantilever array-based sensor devices widely utilise the laser-based optical deflection method for measuring static cantilever deflections mostly with home-built devices with individual geometries. in contrast to scanning probe microscopes,cantilever array devices have no additional positioning device like a piezo stage. as the cantilevers are used in more and more sensitive measurements,it is important to have a simple,rapid,and reliable calibration relating the deflection of the cantilever to the change in position measured by the position-sensitive detector. we present here a simple method for calibrating such systems utilising commercially available afm cantilevers and the equipartition theorem. copyright © 2012 rohit mishra et al.
آدرس centre for research on adaptive nanostructures and nanodevices,school of physics,trinity college dublin, Ireland, institut jacques monod,cnrs,université paris diderot, France, centre for research on adaptive nanostructures and nanodevices,school of physics,trinity college dublin, Ireland
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved