   LC-MS-MS measurements of urinary creatinine and the application of creatinine normalization technique on cotinine in smokers' 24 hour urine  
   
نویسنده hou h. ,xiong w. ,zhang x. ,song d. ,tang g. ,hu q.
منبع journal of analytical methods in chemistry - 2012 - دوره : 2012 - شماره : 0
چکیده    A simple and sensitive high performance liquid chromatography-tandem mass spectrometry (hplc-esi-ms-ms) method was developed and validated for the quantification of creatinine in human urine. the analysis was carried out on an agilent zorbax eclipse xdb-c18 column (2.1×150 mm,3.5 m). the mobile phase was 0.1% formic acid in water and 0.1% formic acid in acetonitrile (50/50,v/v). linear calibration curves were obtained in the concentration range of 1-2000.0 ng/ml,with a lower limit of quantification of 0.99 ng/ml. the intra- and interday precision (rsd) values were below 3%. the method was successfully applied to a bioequivalence study of creatinine in chinese smokers and nonsmokers. the total cotinine in 24 h urine and cotinine: creatinine ratio were also positively associated (pearson r=0.942,p<0.0001). however,cotinine: creatinine ratio varied significantly across smoking groups for the difference of individual. 24 h urinary cotinine was more appropriate for expressing correlation with tar than cotinine: creatinine ratio. © 2012 hongwei hou et al.
آدرس china national tobacco quality supervision,test center, China, china national tobacco quality supervision,test center, China, china national tobacco quality supervision,test center, China, department of urology,first affiliated hospital,zhengzhou university, China, china national tobacco quality supervision,test center, China, china national tobacco quality supervision,test center, China
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved