   the comparison between the impacts of henna and nail polish on pulse oximetry among healthy women  
   
نویسنده moradi majd parisa ,zolfaghari mitra ,behesht aeen fatemeh ,mohseni alireza ,azimi khadijeh ,haghani hamid
منبع nursing practice today - 2014 - دوره : 1 - شماره : 3 - صفحه:120 -125
چکیده    Background & aim: pulse oximetry is a frequently used and standard non-invasive method for monitoring oxygen (o2)-saturation in blood. many factors including dark skin and pigmentation may effect on rate of saturation of the blood oxygen absorbed by pulse oximetry. the effect of nail polish and/or henna color on blood oxygen has not been yet identified and the present study has been carried out by aiming at the review on impact of henna and nail polish on results of pulse oximetry.methods & materials: in the current investigation, clinical trial was studied on 60 resident young women at ages 20-40 by means of purposeful sampling method. initially, 20 g of iranian original henna was solved in 30 ml water and put on forefinger of non-dominant hand of the subjects. theother fingers of the same hand were stained by red, black, and white nail polish, respectively. the middle finger of the same hand was considered as the control variable. then, blood o2-saturation was measured by two calibrated pulse oximetry devices simultaneously.results: the results indicated that henna (p = 0.020), red nail polish (p ? 0.001), and white nail polish (p = 0.020) have increased significantly the rate of o2-saturation absorbed by pulse oximetry. the impact of black nail polish (p = 0.100) on o2-saturation was not significant, but it has changedthe mean rate of o2-saturation. test result of anova with iterative values of f = 10.385 and p ? 0.001 showed the significant statistical difference among mean values of o2-saturation (henna, red, black, white, and control nail polish).conclusion: henna and nail polish may effect on percent of o2-saturation that showed by pulse oximetry and this may lead to error in monitoring of the patient. as a result, it is recommended to use other areas of the hand to put pulse oximetry sensor if henna is utilized and nail polish to be removed before installing the given sensor.
کلیدواژه pulse oximetry ,henna ,nail polish ,blood oxygen saturation
آدرس iran university of medical science, paramedical faculty, department of anesthesia, school of allied medical sciences, iran university of medical science, tehran, iran, ایران, tehran university of medical sciences tums, department of e-learning in medical education, virtual school , tehran university of medical sciences, tehran, iran, ایران, tehran university of medical sciences tums, department of critical care, school of nursing and midwifery, tehran university of medical science, tehran, iran, ایران, mazandaran university of medical sciences, paramedical faculty, department of laboratory sciences, school of allied medical sciences, mazandaran university of medical science, sari, iran, ایران, tehran university of medical sciences tums, department of critical care, school of nursing and midwifery, tehran university of medical science, tehran, iran, ایران, ehran university of medical sciences, department of biostatistics, school of management and information technology, tehran university of medical sciences, tehran, iran, ایران
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved