   improving the image quality of short-time bone spect using cadmium-zinc-telluride detectors with swiftscan  
   
نویسنده funakoshi kazuto ,yamane tomohiko ,kozawa eito ,matsunari ichiro
منبع asia oceania journal of nuclear medicine and biology - 2025 - دوره : 13 - شماره : 1 - صفحه:87 -93
چکیده    Objective(s): this study aimed to evaluate the quality and associated quantitative values of bone single-photon emission computed tomography (spect) with and without swiftscan using a semiconductor camera equipped with a cadmium-zinc-telluride detector.methods: ten patients with bone metastases from prostate cancer who underwent list-mode spect/computed tomography using a whole-body semiconductor camera participated in this study. a total of 130 metastatic lesions from 10 patients were analyzed. standard spect images were obtained approximately 3 h later, and the images were constructed with and without swiftscan.results: the visual assessment of 3-dimensional maximum intensity projection images showed that when an image quality score of 4 (good) or better was considered clinically acceptable, it was maintained at 4 or better in the 75% and 50% scans with swiftscan, whereas only the 75% scan was considered acceptable without swiftscan. the intraclass correlation coefficient was 0.952 at 5% for the standard time without swiftscan and 0.990 with swiftscan. the maximum standardized uptake value (suvmax) changes were 0 to 9.5 (median 1.1) at 75%, 0.1 to 11.5 (1.65) at 50%, 0 to 15.7 (2.1) at 25%, 0.1 to 33.2 (4.2) at 10%, 0.2 to 8.9 (5.65) at 5% without swiftscan. on the contrary, the suvmax changes in absolute value were 0 to 5.4 (median 0.8) at 75%, 0 to 6.5 (1.4) at 50%, 0 to 19.1 (1.7) at 25%, 0 to 24.2 (2.8) at 10%, 0 to 29.9 (2.6) at 5% with swiftscan. the contrast-to-noise ratios (cnr) were 95.3 at 75%, 88.3 at 50%, 69.2 at 25%, 45.7 at 10%, and 31.6 at 5% without swiftscan, and 96.9, 91.7, 78.0, 71.6, and 62.0, respectively, using swiftscan.conclusion: with the use of swiftscan, a 50% reduction in acquisition time was considered acceptable for image quality with reproducible quantitative indices such as suvmax and cnr.
کلیدواژه czt semiconductor camera ,swiftscan ,short-time bone spect/ct
آدرس saitama medical university hospital, department of radiation oncology, japan, kobe city medical center general hospital, department of molecular imaging research, japan. saitama medical university, international medical center, department of nuclear medicine, japan, saitama medical university hospital, department of radiology, japan, saitama medical university hospital, department of nuclear medicine, japan
پست الکترونیکی matsunarii@gmail.com
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved