Fa   |   Ar   |   En
   on the optimization of the protocol for automated radiosyntheses of [68ga]ga-pentixafor, [68ga]ga-fapi-4 and [68ga]ga-dotatate in a modular-lab standard  
   
نویسنده menon sreeja ,mitra arpit ,sahu sudeep ,lad sangita ,chakraborty avik ,ray mukti ,banerjee sharmila
منبع asia oceania journal of nuclear medicine and biology - 2024 - دوره : 12 - شماره : 2 - صفحه:149 -160
چکیده    Objective: the present work describes the automated radiochemical syntheses of different pet tracers like [68ga]ga-pentixafor, [68ga]ga-fapi-4 and [68ga]ga-dotatate using optimized single protocol in the non-cassette based eckert & ziegler (ez) modular lab (fixed tubing system) without any modification in the inbuilt human machine interface (hmi) software. recently, pet agents viz. [68ga]ga-pentixafor and [68ga]ga-fapi-4 are gaining prominence for the diagnosis of overexpressed chemokine receptor-4 (cxcr4) and fibroblast activation protein (fap) receptor, respectively, in the microenvironment of numerous cancer types. the promising results observed with the clinical usage of [68ga]ga-dotatate produced using the automated protocol, provided impetus for the clinical translation of [68ga]ga-pentixafor and [68ga]ga-fapi-4 using the in-house developed automated radiolabeling protocol.methods: herein we report a single radiolabeling protocol for the automated preparation of [68ga]ga-pentixafor and [68ga]ga-fapi-4 in the non-cassette based ez modular-lab standard radiochemistry module, without any changes in schematic, graphical user interface (gui) software and time list, from that used for routine production of [68ga]ga-dotatate in our centre, since 2015. physico-chemical quality control and in-vitro stability analyses were carried out using radio-tlc and radio-hplc.results: the automated protocol yielded reliable and consistent non-decay corrected (ndc) radiochemical yield (rcy) of (84.4±0.9)% and (85.5±1.4)% respectively, for [68ga]ga-pentixafor and [68ga]ga-fapi-4, with rcp> 98%, which are comparable to the rcy of (84.4±1.2)% and rcp (99.1±0.3)% for [68ga]ga-dotatate. the biological quality control studies confirmed the formulations to be of ready-to-use pharmaceutical grade.conclusion: the consistent and reliable rcy and rcp of multiple 68ga-labeled pet tracers by single optimized automated radiochemistry protocol exhibits the versatility of the ez modular lab.
کلیدواژه [68ga]ga-pentixafor ,[68ga]ga-fapi-4 ,[68ga]ga-dotatate ,modular-lab standard ,pet/cta
آدرس bhabha atomic research centre, health physics division, india, board of radiation and isotope technology, radiopharmaceutical laboratory, india, bhabha atomic research centre, radiation medicine centre, india, bhabha atomic research centre, radiation medicine centre, india, homi bhabha national institute, india, homi bhabha national institute, india, homi bhabha national institute, india
پست الکترونیکی sharmila@barc.gov.in
 
     
   