removal of a broken intramedullary nail: a case report and technical description
|
|
|
|
|
|
|
|
نویسنده
|
poutoglidou freideriki ,krkovic matija
|
|
منبع
|
the archives of bone and joint surgery - 2022 - دوره : 10 - شماره : 11 - صفحه:982 -985
|
|
چکیده
|
The removal of a broken intramedullary nail is a challenging procedure. several surgical techniques have beendescribed to remove the distal end of the nail. here, we report the surgical technique for removing broken fitbone®lengthening nail using a cerclage wire. this is an effective and reproducible technique that does not require specializedequipment, the development of a cortical window, or an arthrotomy and is applicable for all types of intramedullary nails.level of evidence: iv
|
|
کلیدواژه
|
broken nail ,cerclage wire ,fitbone nail ,removal
|
|
آدرس
|
cambridge universityhospitals nhs trust, hospitals nhs trust, addenbrookes major trauma unit, department of trauma and orthopaedicsof trauma and orthopaedics, u k, cambridge universityhospitals nhs trust, hospitals nhs trust, addenbrookes major trauma unit, department of trauma and orthopaedics, ukuk
|
|
پست الکترونیکی
|
krkovic@gmail.com
|
|
|
|
|
|
|
|
|
|
|
|
|
-
|
|
|
|
|
Authors
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|