   Hyperglycemia-Induced T-Wave Oversensing as a Cause of Cardiac Resynchronization Therapy (CRT) Failure  
   
نویسنده Assadian Rad Mohammad ,Emkanjoo Zahra ,Moladoust Hassan
منبع journal of tehran university heart center - 2012 - دوره : 7 - شماره : 1 - صفحه:30 -32
چکیده    T-wave oversensing occurs when the counter starts giving dual beeps for every cardiac cycle instead of one. this usually happens when the monitoring lead displays a tall t wave, which is also sharp. r wave sensing algorithms of the devices do not sense t wave because the slow rate of the t wave is much less than that of the r wave. but the slow rate of t waves may change with time and also because of parameters like potassium levels and hyperglycemia. we present a 67-year-old female who underwent the implantation of cardiac resynchronization therapy (cardiac resynchronization and implantable cardioverter defibrilator [crt-d]) because of severe left ventricular systolic dysfunction and ventricular dyssynchrony experienced recurrent inappropriate implantable cardioverter-defibrillator (icd) shocks and crt failure. device analysis showed that the crt failure was in consequence of t-wave oversensing due to hyperglycemia. elimination of the t-wave oversensing after hyperglycemia control conferred good biventricular pacing and good response to crt during a 6-month follow-up period
کلیدواژه Cardiac resynchronization therapy ، Cardiac pacing ,artficial ، Hyperglycemia
آدرس guilan university of medical sciences, Heshmat Cardiovascular Medical Center, ایران, guilan university of medical sciences, Heshmat Cardiovascular Medical Center, ایران, guilan university of medical sciences, Heshmat Cardiovascular Medical Center, ایران
پست الکترونیکی dr_a_asadian@yahoo.com
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved