Hyperglycemia-Induced T-Wave Oversensing as a Cause of Cardiac Resynchronization Therapy (CRT) Failure
|
|
|
|
|
نویسنده
|
Assadian Rad Mohammad ,Emkanjoo Zahra ,Moladoust Hassan
|
منبع
|
journal of tehran university heart center - 2012 - دوره : 7 - شماره : 1 - صفحه:30 -32
|
چکیده
|
T-wave oversensing occurs when the counter starts giving dual beeps for every cardiac cycle instead of one. this usually happens when the monitoring lead displays a tall t wave, which is also sharp. r wave sensing algorithms of the devices do not sense t wave because the slow rate of the t wave is much less than that of the r wave. but the slow rate of t waves may change with time and also because of parameters like potassium levels and hyperglycemia. we present a 67-year-old female who underwent the implantation of cardiac resynchronization therapy (cardiac resynchronization and implantable cardioverter defibrilator [crt-d]) because of severe left ventricular systolic dysfunction and ventricular dyssynchrony experienced recurrent inappropriate implantable cardioverter-defibrillator (icd) shocks and crt failure. device analysis showed that the crt failure was in consequence of t-wave oversensing due to hyperglycemia. elimination of the t-wave oversensing after hyperglycemia control conferred good biventricular pacing and good response to crt during a 6-month follow-up period
|
کلیدواژه
|
Cardiac resynchronization therapy ، Cardiac pacing ,artficial ، Hyperglycemia
|
آدرس
|
guilan university of medical sciences, Heshmat Cardiovascular Medical Center, ایران, guilan university of medical sciences, Heshmat Cardiovascular Medical Center, ایران, guilan university of medical sciences, Heshmat Cardiovascular Medical Center, ایران
|
پست الکترونیکی
|
dr_a_asadian@yahoo.com
|
|
|
|
|
|
|
|
|
|
|
|
Authors
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|