INFLUENCE OF THE MEASUREMENT MEDIUM AND MATRIX MODIFIERS ON THE DETERMINATION OF SILICON IN WATERS BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROMETRY
|
|
|
|
|
نویسنده
|
EL HIMRI MAMOUNE ,EL HIMRI ABDELOUAHAD ,PASTOR AGUSTÍN ,DE LA GUARDIA MIGUEL
|
منبع
|
journal of applied sciences and environmental management - 2013 - دوره : 17 - شماره : 1 - صفحه:37 -41
|
چکیده
|
An electrothermal atomic absorption (etaas) method has been developed for the direct determination of si in natural waters. measurements were carried out on previously alkalinized samples and using a mixture of pd and mg(no3)2 as matrix modifier, with an ashing temperature of 1400 °c and an atomization temperature of 2400 °c. transversally heated pyrolitically graphite tubes, equipped with l’vov platform and zeeman effect correction, was employed in this study. a limit of detection of 2.6 μg/l si was obtained and accuracy was evaluated by analyzing two standard reference material at different concentration levels, a good agreement (<5%) was obtained. this developed procedure was demonstrated of the need of using an alkaline medium for accurate si determination by etaas.
|
کلیدواژه
|
Silicon determination ,water analysis electrothermal atomic absorption spectrometry ,effect of sample medium ,effet of matrix modifier
|
آدرس
|
University Mohammed premier, Laboratory of chemistry and environnement, Morocco, Université Mohammed Premier, Ecole Nationale des Sciences Appliquées, Morocco, University of Valencia, department of analytical chemistry, Spain, University of Valencia, department of analytical chemistry, Spain
|
|
|
|
|
|
|
|
|
|
|
|
|
|