Fa   |   Ar   |   En
   Using potentiometry and electrochemical impedance spectroscopy techniques for studying effect of nanomaterials on salicylate ion-selective electrode response  
   
نویسنده jahromi z.a.y. ,mazloum-ardakani m. ,zare-mehrjardi h.r. ,vafazadeh r.
منبع analytical and bioanalytical electrochemistry - 2017 - دوره : 9 - شماره : 5 - صفحه:562 -573
چکیده    A novel anion-selective electrode for salicylate ion with new ionophore was designed based on modified carbon paste electrode with multi walled carbon nanotubes (mwcnts). electrode made was evaluated for electrochemical studies using potentiometric and impedance techniques and results compared with together. the results of the potentiometric method using modified carbon paste electrode showed that the prepared electrode has a nernstian slope 60.1±0.1 mv in concentrations measurement range 1.0×10-7-1.0×10-1 m,the detection limit 5.4×10-8,with ph range of 3.5-9.3,while using the impedance technique,the concentration measurement range increased to 1.0×10-9-1.0×10-1 m,ph range increased to 3.8-10.7 and the detection limit decreased to 4.3×10-10 m. © 2017 by cee (center of excellence in electrochemistry).
کلیدواژه Chemometrics; Electrochemical impedance spectroscopy; Multi-walled carbon nanotube; Nanomaterials; Potentiometry
آدرس department of chemistry,payame noor university,p.o. box 19395-4697, ایران, department of chemistry,faculty of science,yazd university, ایران, department of chemistry,payame noor university,p.o. box 19395-4697, ایران, department of chemistry,faculty of science,yazd university, ایران
 
     
   