Fa   |   Ar   |   En
   Development and Validation of a New Capillary Electrophoresis Assay for Monitoring of Atenolol Photodegradation Product in Non-aqueous Solution  
   
نویسنده Hasanzadeh Mohammad ,Shadjou Nasrin ,Kazeman Isa ,Jouyban Abolghasem
منبع analytical and bioanalytical electrochemistry - 2014 - دوره : 6 - شماره : 2 - صفحه:219 -233
چکیده    The characterization of atenolol photodegradation product was performed using capillary electrophoresis (ce) with ultraviolet diode array detector (uv-dad). the optimum separation for this assay was achieved in < 7 min at 298 k with a fused-silica capillary column (57 cm length and 75 ?m i.d.) and a running buffer containing 60 mm acetate buffer at ph 5.3 dissolved in methanol/ethanol mixture (20:80% v/v). the samples were injected with applied voltage of 16.0 kv. it’s established that capillary electrophoresis method enables to detect atenolol photodegradation product. the suggested method has been used for the determination of the studied drug in human serum and the results were compared to the other reported methods
کلیدواژه Capillary electrophoresis (CE) ,Atenolol ,Ultraviolet diode array detector (UV-DAD) ,Photodegradation
آدرس tabriz university of medical sciences, Drug Applied Research Center and Faculty of Pharmacy, Tabriz University of Medical Sciences, Tabriz 51664, Iran, ایران, Pasteur institute of Iran, Department of biochemistry, Pasteur institute of Iran, Tehran, Iran, ایران, k.n.toosi university of technology, Department of Chemistry, K N Toosi University of Technology, Tehran, Iran, ایران, tabriz university of medical sciences, Drug Applied Research Center and Faculty of Pharmacy, Tabriz University of Medical Sciences, Tabriz 51664, Iran,, ایران
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved