Electrochemical Determination of Antibacterial Drug Moxifloxacin Hydrochloride Using Chloranil Modified Carbon Paste Electrode
|
|
|
|
|
|
|
|
نویسنده
|
Attia Ali K. ,El-Shal Manal A.
|
|
منبع
|
analytical and bioanalytical electrochemistry - 2012 - دوره : 4 - شماره : 3 - صفحه:213 -224
|
|
چکیده
|
A highly sensitive electroanalytical method for the determination of moxifloxacinhydrochloride using differential pulse voltammetry was described. the electrochemicalbehavior of the used drug was investigated in britton-robinson buffer in the ph range of 2.0to 11.0 at chloranil modified carbon paste electrode. the anodic peak currents increasedlinearly with the concentration in the range of 4.0×10-7 to 3.6×10-6 mol l-1. the limits ofdetection and quantification were 6.74×10-8 mol l-1 and 2.25×10-7 mol l-1, respectively. therelative standard deviation was found in the following range: 0.569-0.974%. the percentagerecovery was found in the following range: 99.28-100.60%. the method was successfullyapplied to the determination of moxifloxacin hydrochloride in tablets without previousseparation.
|
|
کلیدواژه
|
Moxifloxacin Hydrochloride ,Chloranil ,Oxidation ,Differential Pulse Voltammetry ,Tablets
|
|
آدرس
|
Cairo university, National Organization for Drug Control and Research, P O Box 29, Cairo, Egypt, Egypt, Cairo university, National Organization for Drug Control and Research, P O Box 29, Cairo, Egypt, ایران
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|