This study describes the preparation and application of a poly(3,4-ethylenedioxythiophene) (pedot) modified glassy carbon electrode for paracetamoldetermination in pharmaceutical samples. the pedot modified electrode was preparedby electropolymerizing 3,4-ethylenedioxythiophene (edot) on a glassy carbon electrodein a non-aqueous medium. the best performance of the pedot modified electrode in 0.1mol l-1 phosphate buffer obtained at ph 7.0. under these conditions, an oxidationpotential of paracetamol was observed at 0.37 v versus ag/agcl/sat’d kcl. the cyclicvoltammetric study indicates that the pedot modified electrode shows a very goodelectrocatalytic activity by reducing the overpotential by 200 mv for paracetamoloxidation. the modified electrode showed a linear response range between 2.5 and 150?mol l-1, a sensitivity of 90.03 ?a/mmol l-1, quantification and detection limits of 2.5?mol l-1 and 1.13 ?mol l-1, respectively. finally, the proposed method was applied forparacetamol determination in commercial tablets with a mean recovery of (101?6) %.
|
|
کلیدواژه
|
Paracetamol ,PEDOT ,Glassy Carbon (GC) ,Differential Pulse Voltammetry
|
|
آدرس
|
Department of Chemistry, Addis Ababa University, P.O. Box 1176, Addis Ababa, Ethiopia, Ethiopia, Department of Chemistry, Addis Ababa University, P.O. Box 1176, Addis Ababa, Ethiopia, Ethiopia, Department of Chemistry, Addis Ababa University, P.O. Box 1176, Addis Ababa, Ethiopia, Ethiopia, Department of Chemistry, Addis Ababa University, P.O. Box 1176, Addis Ababa, Ethiopia, Ethiopia
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Authors
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|