Evaluation of the Usefulness of an Electrochemical Sensor in Detecting Ascorbic Acid using a Graphite Screen-printed Electrode Modified with NiFe_2O_4 Nanoparticles
|
|
|
|
|
|
|
|
نویسنده
|
jahani shohreh
|
|
منبع
|
analytical and bioanalytical electrochemistry - 2018 - دوره : 10 - شماره : 6 - صفحه:739 -750
|
|
چکیده
|
In this study, a new modified electrode was proposed to measure ascorbic acid. graphite screen printed electrode was modified with nife_2o_4 nanoparticles (nife-2o_4/spe). cyclic voltammetry (cv), chronoamperometry (cha) and differential pulse voltammetry (dpv) methods were used to analyze the prepared electrode. the catalytic property resulting from the synergistic effect of this modifier not only increased the oxidation current but also reduced the oxidation potential of ascorbic acid. scan rate and ph effects on the performance of nife_2o_4/spe were also investigated. the linear calibration curve of ascorbic acid was obtained within the linear range of 5.0×10^−7– 1.0×10^−4 m, and the detection limit was calculated as 1.0×10^−7 m. finally, the modified electrode was successfully used to measure ascorbic acid in some real samples.
|
|
کلیدواژه
|
Ascorbic acid ,NiFe_2O_4 nanoparticles ,Voltammetric sensor ,Screen printed electrode
|
|
آدرس
|
islamic azad university, kerman branch, young researchers and elite club, Iran
|
|
پست الکترونیکی
|
shohre_jahani@yahoo.com
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|