   Robust control of LVAD based on the sub-regional modeling of the heart  
   
نویسنده ravanshadi s. ,jahed m.
منبع scientia iranica - 2016 - دوره : 23 - شماره : 6-D - صفحه:2934 -2943
چکیده    Left ventricular assist devices (lvad) have received renewed interest as a bridge-to-transplantation as well as a bridge-to-recovery device. ironically, reports of malfunction and complications have hindered the growth of this device. in particular, the main concern is lvad's susceptibility to excessive backlash and suction as a result of ows that are either too low or high, respectively. this study utilizes a wellestablished physiological model of the cardiovascular system as a reliable platform to study a proposed adaptive robust controller for a rotary motor based lvad which overcomes such shortcomings. proposed controller performance is evaluated by comparing simulated natural heart model with lvad assisted diseased heart in various states, extending from 60 to 130 beats per minute (bpm). simulation results of the proposed lvad controller show that for heart rate of 75 bpm, systolic and diastolic blood pressures are 112 ± 18 mmhg and 73±16 mmhg, respectively. furthermore, for the light exercise condition of 130 bpm, systolic and diastolic blood pressures increase to 155 ± 19 mmhg and 96 ± 14 mmhg, respectively. these results closely match natural heart clinical measurements, con rming proposed lvad model and its adaptive robust controller to be a possible solution to currentissues confronting the lvad drives.
کلیدواژه Ventricular assist device; Adaptive robust control; Distributed model; Cardiovascular system.
آدرس sharif university of technology, school of electrical engineering, ایران, sharif university of technology, school of electrical engineering, ایران
پست الکترونیکی jahed@sharif.edu
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved