   Tradeoffs between Radiation Exposure to the Lens of the Eyes and Diagnostic Image Quality in Pediatric Brain Computed Tomography  
   
نویسنده karami vahid ,albosof mohsen ,gholami mehrdad ,hekmatnia ali ,fallah bagher sheidaei mehdi ,taghizadeh behbahani ali ,sadat sharif hoda ,jafrasteh somayeh ,adeli mohammad
منبع journal of medical signals and sensors - 2023 - دوره : 13 - شماره : 3 - صفحه:208 -216
چکیده    Background: computed tomography (ct) of the brain is associated with radiation exposure to the lens of the eyes. therefore, it is necessary to optimize scan settings to keep radiation exposure as low as reasonably achievable without compromising diagnostic image information. the aim of this study was to compare the effectiveness of the five practical techniques for lowering eye radiation exposure and their effects on diagnostic image quality in pediatric brain ct. methods: the following scan protocols were performed: reference scan, 0.06‑mm pbeq bismuth shield, 30% globally lowering tube current (gltc), reducing tube voltage (rtv) from 120 to 90 kvp, gantry tilting, and combination of gantry tilting with bismuth shielding. radiation measurements were performed using thermoluminescence dosimeters. objective and subjective image quality was evaluated. results: all strategies significantly reduced eye dose, and increased the posterior fossa artifact index and the temporal lobe artifact index, relative to the reference scan. gltc and rtv increased image noise, leading to a decrease signal‑to‑noise ratio and contrast‑to‑noise ratio. except for bismuth shielding, subjective image quality was relatively the same as the reference scan. conclusions: gantry tilting may be the most effective method for reducing eye radiation exposure in pediatric brain ct. whenthe scanner does not support gantry tilting, gltc might be an alternative.
کلیدواژه Brain computed tomography ,eye lens ,image quality ,radiation exposure
آدرس dezful university of medical sciences, ganjavian hospital, clinical research development unit, Iran, dezful branch, islamic azad university, department of biomedical engineering, Iran, lorestan university of medical sciences, department of medical physics, Iran, isfahan university of medical sciences, department of radiology, Iran, dezful university of medical sciences, ganjavian hospital, clinical research development unit, Iran, islamic azad university, dezful branch, department of biomedical engineering, Iran, dezful university of medical sciences, ganjavian hospital, clinical research development unit, Iran, dezful university of medical sciences, ganjavian hospital, clinical research development unit, Iran, islamic azad university, dezful branch, department of biomedical engineering, Iran
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved