Fa   |   Ar   |   En
   The Design of an Audit Test for 60Co Brachytherapy Treatment Planning System  
   
نویسنده tahmasebi birgani mohammad javad ,khorshidsavar hadis ,bagheri ali ,danyaei amir ,abdalvand neda
منبع journal of medical signals and sensors - 2022 - دوره : 12 - شماره : 2 - صفحه:133 -137
چکیده    Background: auditing the treatment planning system (tps) software for a radiotherapy unit is of paramount importance in any radiation therapy department. a plexiglas phantom was proposed to measure the ionization of 60co high dose rate (hdr) source and compare dose points in the planning system for auditing and verifying tps. methods: auditing was performed using a plexiglas phantom in an end‑to‑end test, and relative dose points were detected by a farmer‑type ionization chamber and compared with the relative dose of similar points in tps. the audit results were determined as pass optimal level (<3.3%), pass action level (between 3.3% and 5%), and out of tolerance (>5%). results: the comparison of the collected data revealed that 80% of the measured values were ≤5% in the pass level, and 20% of the points were out of tolerance (between 5% and 6.99%). conclusion: this study documented the appropriateness of the dosimetry audit test and this phantom design for the hdr brachytherapy tps.
کلیدواژه Audit ,brachytherapy ,dosimetry ,Plexiglas phantom ,radiotherapy
آدرس ahvaz jundishapur university of medical science, medicine school, golestan hospital, department of radiation oncology, department of medical physics, Iran, ahvaz jundishapur university of medical science, medicine school, department of medical physics, Iran, ahvaz jundishapur university of medical science, golestan hospital, medicine school, department of radiation oncology, Iran, ahvaz jundishapur university of medical science, medicine school, department of medical physics, Iran, ahvaz jundishapur university of medical science, golestan hospital, department of radiation oncology, Iran. iran university of medical science, school of medicine, department of medical physics, Iran
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved