   A Comparison of Dose Distribution Parameters Between Conventional and Mono Isocentric Methods of Breast Irradiation in Phantom By Gafchromic Ebt2 Films  
   
نویسنده
منبع Basic And Clinical Cancer Research - 2014 - دوره : 6 - شماره : 2 - صفحه:35 -41
چکیده    Background: this study was set to determine the optimal technique of breast cancer with supraclavicular fossa irradiation. methods: three techniques were compared in this study. techniques were two point symmetric (2p-sym), two point asymmetric (2p-asym) and single point (sp). in 2p-sym all of beams were symmetric, in 2p-asym tangential and supraclavicular beams respectively were symmetric and asymmetric, in sp all of beams were asymmetric. in this studyhas been done dosimetry by gafchromic films in phantom with compact accelerator. techniqueswere analyzed by dvhs treatment’s volumes and organ at risk (oar) volumes. results: there are no notable differences between dose distributions in three techniques but hot spot in the junction of treatment’s beams in single point (sp) technique was 115% and 118% for other techniques. the analysis of dvhs showed a decrease in oar’s doses with sp technique. there was a significant difference of v20%, v20, v30 for lungs and v10, v40 for heart in sp technique compared to other techniques. the profiles of dose that obtained from films showed a gap in matchline, while matchline’s dose distribution in software was right and there was no gap. discussion: observed gap in gafchromic films was created because of mechanical limitations in accelerators. correction methods for this limitation were offered in many articles. conclusion: sp technique is optimal technique compared to other techniques if correction methods are done. it seems that iranian radiotherapy clinics don’t pay attention to this correction methods.
کلیدواژه Breast Cancer ,Mono Isocentric ,Phantom ,Gafchromic Ebt2
آدرس
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved