quantification of artifacts and image distortions in 1.5 tesla magnetic resonance images of a commercial multi-channel vaginal cylinder brachytherapy applicator set
|
|
|
|
|
نویسنده
|
kanani abolfazl ,fatemi-ardakani ali ,owrangi amir m. ,yazdi mehran ,baghbani hadi ,mosleh-shirazi mohammad amin
|
منبع
|
journal of biomedical physics and engineering - 2023 - دوره : 13 - شماره : 6 - صفحه:523 -534
|
چکیده
|
Background: the bebig portio multi-channel applicator provides better target dose coverage and sparing organs-at-risk compared to a single-channel cylinder. however, artifacts and distortions of portio in magnetic resonance images (mri) have not yet been reported.objective: we aimed to quantify the artifacts and distortions in its 1.5-tesla mr images before clinical use.material and methods: in this experimental study, we employed a gelatin-filled phantom to conduct our measurements. t2-weighted (t2w) images were examined for artifacts and distortions. computed tomography (ct) images were used as a reference to assess image distortions. artifact severity was measured by recording the full-width-at-half-maximum (fwhm) image pixel values at various positions along the length of the applicator/channels. ct and mri-based applicator reconstruction accuracy were then compared, and signal-to-noise ratio (snr) and contrast were also determined for the applicator images.results: the applicator distortion level for the portio applicator was less than the image spatial resolution (0.5±0.5 pixels). the average fwhm for the tandem applicator images was 5.23±0.39 mm, while it was 3.21±0.37 mm for all channels (compared to their actual diameters of 5.0 mm and 3.0 mm, respectively). the average applicator reconstruction difference between ct and mr images was 0.75±0.30 mm overall source dwell positions. the image snr and contrast were both acceptable. conclusion: these findings indicate that the portio applicator has a satisfactory low level of artifacts and image distortions in 1.5-tesla, t2w images. it may, therefore, be a promising option for mri-guided multi-channel vaginal brachytherapy.
|
کلیدواژه
|
magnetic resonance imaging ,radiotherapy ,brachytherapy ,portio applicator ,image processing ,gynecolgical ,endometrial cancer
|
آدرس
|
shiraz university of medical sciences, ionizing and non-ionizing radiation protection research center (inirprc), school of paramedical sciences, iran, jackson state university (jsu), department of physics, usa, ut southwestern medical center, department of radiation oncology, usa, shiraz university, school of electrical and computer eng, signal and image processing lab (sipl), iran, shiraz university of medical sciences, namazi hospital, department of radiology, iran, shiraz university of medical sciences, ionizing and non-ionizing radiation protection research center (inirprc), school of paramedical sciences, iran
|
پست الکترونیکی
|
mosleh_amin@hotmail.com
|
|
|
|
|
|
|
|
|
|
|
|
Authors
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|