Fa   |   Ar   |   En
   conceptual design of a device for online calibration of spirometer based on neural network  
   
نویسنده torbati nasim ,nafisi vahid reza ,eghbal manouchehr
منبع journal of biomedical physics and engineering - 2023 - دوره : 13 - شماره : 3 - صفحه:291 -296
چکیده    Daily calibration of spirometry devices plays an important role in promoting the accuracy of pulmonary diagnostic results. it is needed to have more precise and adequate instruments for calibrating spirometry during the clinical use. in this work, a device was designed and developed based on a calibrated-volume syringe and an electrical circuit was also built to measure the air flux. some colored tapes with specific size and order covered the syringe piston. when the piston moved in front of the color sensor, the input air flow was calculated according to the width of the strips and transferred to the computer. a radial basis function (rbf) neural network estimator used new data to modify the previous estimation function for increasing the accuracy and the reliability. the simulation showed that the root mean square of the error improved from 13.7±0.37% to 4.2±0.22%, i.e. the calibration curve has improved about 70%.
کلیدواژه spirometry; calibration; artificial intelligence; neural network model; color sensor
آدرس ministry of health and medical education, food and drug administration, iran, iranian research organization for science and technology, biomedical engineering group, department of electrical & information technology, iran, iranian research organization for science and technology, biomedical engineering group, department of electrical & information technology, iran
پست الکترونیکی eghbal_m@yahoo.com
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved