assessment of surface and build-up doses for a 6 mv photon beam using parallel plate chamber, ebt3 gafchromic films, and primo monte carlo simulation code
|
|
|
|
|
نویسنده
|
mahur mamta ,singh munendra ,gurjar om prakash ,semwal manoj kumar
|
منبع
|
journal of biomedical physics and engineering - 2022 - دوره : 12 - شماره : 5 - صفحه:455 -464
|
چکیده
|
Background: accurate assessment of surface and build-up doses has a key role in radiotherapy, especially for the superficial lesions with uncertainties involved while performing measurements in the build-up region. objective: this study aimed to assess surface and build-up doses for 6 mv photon beam from linear accelerator using parallel plate ionization chamber, ebt3 gafchromic films, and primo monte carlo (mc) simulation code. material and methods: in this experimental study, parallel plate chamber (ppc05) and ebt3 gafchromic films were used to measure doses in a build-up region for 6 mv beam from the linear accelerator for different field sizes at various depths ranging from 0 to 2 cm from the surface with 100 cm source to surface distance (ssd) in a solid water phantom. measured results were compared with monte carlo simulated results using penelope-based primo simulation code for the same setup conditions. effect of gantry angle incidence and ssd were also analyzed for depth doses at the surface and build-up regions using ppc05 ion chamber and ebt3 gafchromic films. results: doses measured at the surface were 14.78%, 19.87%, 25.83%, and 31.54% for field sizes of 5×5, 10×10, 15×15, and 20×20 cm², respectively for a 6 mv photon beam with a parallel plate chamber and 14.20%, 19.14%, 25.149%, and 30.90%, respectively for ebt3 gafchromic films. both measurement sets were in good agreement with corresponding simulated results from the primo mc simulation code; doses increase with the increase in field sizes. conclusion: good agreement was observed between the measured depth doses using parallel plate ionization chamber, ebt3 gafchromic films, and the simulated depth doses using primo monte carlo simulation code.
|
کلیدواژه
|
chamber ,dosimetry ,monte carlo method ,radiotherapy
|
آدرس
|
delhi state cancer institute, department of radiation oncology, india. sharda university, school of basic sciences and research, department of physics, india, sharda university, school of basic sciences and research, department of physics, india, mahatma gandhi memorial medical college, government cancer hospital, india, army hospital (research & referral), department of radiation oncology, india
|
پست الکترونیکی
|
manojsemwal@yahoo.co.in
|
|
|
|
|
|
|
|
|
|
|
|