Fa   |   Ar   |   En
   impact of roi size on the accuracy of noise measurement in ct on computational and acr phantoms  
   
نویسنده anam choirul ,triadyaksa pandji ,naufal ariij ,arifin zaenal ,muhlisin zaenul ,setiawati evi ,budi wahyu setia
منبع journal of biomedical physics and engineering - 2022 - دوره : 12 - شماره : 4 - صفحه:359 -368
چکیده    Background: the effect of region of interest (roi) size variation on producing accurate noise levels is not yet studied. objective: this study aimed to evaluate the influence of roi sizes on the accuracy of noise measurement in computed tomography (ct) by using images of a computational and american college of radiology (acr) phantoms. material and methods: in this experimental study, two phantoms were used, including computational and acr phantoms. a computational phantom was developed by using matlab r215a software (mathworks inc., natick, ma natick, ma) with a homogeneously +100 hounsfield unit (hu) value and an added-gaussian noise with various levels of 5, 10, 25, 50, 75, and 100 hu. the acr phantom was scanned with a philips mx-16 slice ct scanner in different slice thicknesses of 1.5, 3, 5, and 7 mm to obtain noise variation. noise measurement was conducted at the center of the phantom images and four locations close to the edge of the phantom images using different roi sizes from 3×3 to 41×41 pixels, with an increased size of 2×2 pixels. results: the use of a minimum roi size of 21×21 pixels shows noise in the range of ±5% ground truth noise. the measured noise increases above the ±5% range if the used roi is smaller than 21×21 pixels. conclusion: a minimum acceptable roi size is required to maintain the accuracy of noise measurement with a size of 21×21 pixels.
کلیدواژه acr phantom; computational phantom; diagnostic imaging; image quality; noise measurement; radiologic phantoms; tomography ,x-ray computed; x-rays
آدرس diponegoro university, faculty of sciences and mathematics, department of physics, indonesia, diponegoro university, faculty of sciences and mathematics, department of physics, indonesia, diponegoro university, faculty of sciences and mathematics, department of physics, indonesia, diponegoro university, faculty of sciences and mathematics, department of physics, indonesia, diponegoro university, faculty of sciences and mathematics, department of physics, indonesia, diponegoro university, faculty of sciences and mathematics, department of physics, indonesia, diponegoro university, faculty of sciences and mathematics, department of physics, indonesia
پست الکترونیکی wahyu.sb@fisika.fsm.undip.ac.id
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved