Fa   |   Ar   |   En
   Performances of evaporative light scattering detector for unmodified cyclodextrins analysis by liquid chromatography  
   
نویسنده El Haddad M. ,Slimani R. ,Mamouni R. ,El Antri S. ,Lazar S.
منبع journal of materials and environmental science - 2010 - دوره : 1 - شماره : 3 - صفحه:151 -156
چکیده    Evaporative light scattering detector (elsd) was used as detection system for isocratic analysis of α-, β-and γ- cyclodextrins by liquid chromatography (lc) on zorbax phenyl (250 x 4.6 mm i. d.) column. an evaluation of elsd response by study of the variation of drift tube temperature was realized. simultaneous analysis of minoritary compound in presence of a majoritary compound without interferences and saturation is investigated by elsd offering the possibility of gain variation.
کلیدواژه Evaporative light scattering detector ,drift tube temperature ,gain variation ,simultaneous analysis ,α- ,β-and γ-cyclodextrins ,liquid chromatography
آدرس Université Hassan II, FST, Laboratoire de Biochimie, Environnement & Agroalimentaire, Morocco. Université Cadi Ayyad, Faculté Poly disciplinaire, Département Sciences de la Matière, Morocco, Université Hassan II, FST, Laboratoire de Biochimie, Environnement & Agroalimentaire, Morocco, Université Hassan II, FST, Laboratoire de Biochimie, Environnement & Agroalimentaire, Morocco, Université Hassan II, FST, Laboratoire de Biochimie, Environnement & Agroalimentaire, Morocco