Fa   |   Ar   |   En
   Monte Carlo modelling of a NaI(Tl) scintillator detectors using MCNP simulation code  
   
نویسنده mouhti i. ,elanique a. ,messous m.y.
منبع journal of materials and environmental science - 2017 - دوره : 8 - شماره : 12 - صفحه:4560 -4565
چکیده    Due to their attractive performances,inorganic scintillators such as nai(tl) detectors are widely used in the field of gamma-ray spectrometry where knowledge of detector efficiency is a key factor for successful quantitative γ ray analysis. monte carlo simulations using mcnp code have been performed in order to evaluate absolute efficiency of nai detectors. two mathematical models were constructed for the nai crystal: 7.62cm x 7.62cm cylindrical nai and 7.62cm diameter spherical nai. detector efficiency was calculated for both nai detectors in the photon energy range from 60 kev to 2750 kev. simulation results were compared to experimental efficiency data. © 2017,university of mohammed premier oujda morocco.
کلیدواژه Absolute efficiency; Gamma-ray spectrometry; MCNP code; Monte Carlo simulation; NaI detector
آدرس faculty of sciences,laboratory of condensed matter physics and nanomaterials for renewable energy,university ibn zohr,bp 8106,agadir, Morocco, faculty of sciences,laboratory of condensed matter physics and nanomaterials for renewable energy,university ibn zohr,bp 8106,agadir, Morocco, national center for energy,sciences and nuclear techniques,cnesten-,b.p 1382,rabat, Morocco
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved