Fa   |   Ar   |   En
   Determination of aluminum in dialysis concentrates by atomic absorption spectrometry after coprecipitation with lanthanum phosphate  
   
نویسنده selvi e.k. ,şahin u. ,şahan s.
منبع iranian journal of pharmaceutical research - 2017 - دوره : 16 - شماره : 3 - صفحه:1030 -1038
چکیده    This method was developed for the determination of trace amounts of aluminum(iii) in dialysis concentrates using atomic absorption spectrometry after coprecipitation with lanthanum phosphate. the analytical parameters that influenced the quantitative coprecipitation of analyte including amount of lanthanum,amount of phosfate,ph and duration time were optimized. the % recoveries of the analyte ion were in the range of 95-105 % with limit of detection (3s) of 0.5 µg l-1. preconcentration factor was found as 1000 and relative standard deviation (rsd) % value obtained from model solutions was 2.5% for 0.02 mg l-1. the accuracy of the method was evaluated with standard reference material (cww-tmd waste water). the method was also applied to most concentrated acidic and basic dialysis concentrates with satisfactory results. © 2017 by school of pharmacy shaheed beheshti university of medical sciences and health services.
کلیدواژه Aluminum; Atomic absorption spectrometer; Coprecipitation; Dialysis concentrate; Lanthanium phosphate
آدرس recep tayyip erdoğan university,faculty of art and science,chemistry department,rize, Turkey, erciyes university,faculty of science,chemistry department,kayseri, Turkey, erciyes university,faculty of agriculture,department of soil science and plant nutrition,kayseri, Turkey
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved