   evaluating diagnostic efficacy of cone beam computed tomography performed in different imaging conditions based on field of view and resolution in the detection of mandibular condyle erosions: an in vitro study  
   
نویسنده sharifi shooshtari sanaz ,razavi mahshid ,boostani fard parisa ,rohani ali ,hamidiaval shadi ,niroomand naghme
منبع iranian journal of radiology - 2019 - دوره : 16 - شماره : 3 - صفحه:1 -8
چکیده    Background:new cone beam computed tomography (cbct) devices have the ability to take images with different fields of views (fovs) and resolutions. the larger the fov and the higher resolution of the image, the higher the patient's dose would be.objectives:this research aims to analyze the diagnostic accuracy of fov's different dimensions and the various resolutions in the diagnosis of mandibular condyle erosions.methods:cbct images using newtom vgi (verona, italy) system in five different fovs and resolutions (voxel size) were taken from eight human dried mandibles in which the erosionlike lesions were created on both condyles. imaging was performed before and after the formation of erosion; afterwards, the images were evaluated by two maxillofacial radiologists to analyze the absence or presence of the lesions. resultant data was evaluated by spss v. 22.0, mcnemar and kappa statistical methods.results:the highest sensitivity, specificity, positive predictive value (ppv), negative predictive value (npv), and accuracy were attributed to 8 × 8 fov and the lowest was attributed to 12 × 8 fov. in analyzing different resolutions, the highest sensitivity, specificity, ppv, and npv were attributed to high resolution (0.125 and 0.15 mm voxel size). there was no statistically significant difference between sensitivity, specificity, ppv, and npv of different fovs and different resolutions. interobserver and intraobserver cofficiency were at excellent range.conclusion:considering there was no significant difference between the results, it is possible to use voxels in larger size to reduce patient's dose in order to detect mandibular condyle erosions. in addition, fov with lower patient's dose is the method of choice.
کلیدواژه mandibular condyle erosions ,cone beam computed tomography (cbct) ,voxel size ,field of view (fov)
آدرس ahvaz jundi shapur university of medical sciences, school of dentistry, department of oral and maxillofacial radiology, iran, ahvaz jundi shapur university of medical sciences, school of dentistry, department of oral and maxillofacial radiology, iran, ahvaz jundi shapur university of medical sciences, school of dentistry, iran, ahvaz jundi shapur university of medical sciences, school of dentistry, iran, lorestan university of medical sciences, department of oral and maxillofacial radiology, iran, arak university of medical sciences, department of oral and maxillofacial radiology, iran
پست الکترونیکی naghmeh8866@gmail.com
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved