   Dosimetric investigation of high dose rate Ir-192 source with Monte Carlo method  
   
نویسنده acun h. ,bozkurt a. ,kemikler g.
منبع international journal of radiation research - 2017 - دوره : 15 - شماره : 3 - صفحه:241 -249
چکیده    Background: this study aims to calculate the air-kerma strength (sk),the dose rate constant (∧) and the dose rate profiles of a gammamed 12i ir-192 source by using the monte carlo technique and to compare the dose rate values with those calculated by the abacus hdr treatment planning system (tps). materials and methods: air-kerma strength (in units of u; 1 u= μgy m2 h-1) and the dose rate constant (∧) of a gammamed 12i ir-192 source were computed with the monte carlo-based code mcnp. for a single dwell posi4on,profiles of dose rate per air kerma rate strength (in unit of u) along the x axis were obtained with mcnp and compared with data from the abacus tps at y=0,1,and 2 cm. results: the air-kerma strength and the dose rate constant of the source were calculated to 9.98×10-8 u and 1.106 cgy h-1 u-1,respec4vely. the maximum dose differences between mcnp and tps along the x axis were found to be 1.3,3.7 and 5.4% at distances more than 1 cm away from the source center for y= 0,1 and 2 cm planes,respec4vely. conclusion: the dose rate profiles calculated with mcnp and by the tps show good agreement except for points located beyond the 4p of the source.
کلیدواژه High dose rate brachytherapy; Ir-192; Monte Carlo
آدرس harran university,faculty of medicine,department of biophysics,osmanbey campus,sanliurfa, Turkey, akdeniz university,faculty of engineering,department of biomedical engineering,antalya, Turkey, istanbul university,institute of oncology,department of medical physics,capa,istanbul, Turkey
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved