   Determination of room entry times for radiation therapists after routine 15 MV photon treatments  
   
نویسنده Najafi M. ,Deevband M.R. ,Yousefi Diba A.A. ,Amin Moghaddam A.
منبع international journal of radiation research - 2015 - دوره : 13 - شماره : 4 - صفحه:379 -382
چکیده    Background: radiation therapy uses high-energy radiation to kill cancer cells.photoneutron contamination and induced radioactivity of high energytherapeutic photon beams are considered as the main source of occupaonalexposure to radiation therapists who works with linear accelerators operating above 10 mv. materials and methods: the gamma dose rates were measured after termination of different treatment approaches using 15 mv photons and room entry times for each approach was determined. based on the results of this study, the annual dose of radiation therapists was estimated. results: the highest dose rate that measured in the treatment room was belonged to 3-field pelvic approach which was equal to 5 ± 1.1 μsv/h. the radiation therapist's room entry times were determined between 3 to 10 minutes and the annual dose was estimated up to 5.9 msv/y. conclusion: although the estimated annual dose is less than the internationally permi2ed value, the undesirable dose to radiation workers could be reduced through considering recommended room entry times.
کلیدواژه Linear accelerator ,high energy photons ,radiation therapists ,room entry times
آدرس shahid beheshti university of medical sciences, Department of Bioengineering and Medical Physics, ایران, shahid beheshti university of medical sciences, Department of Bioengineering and Medical Physics, ایران, Hafte Tir Hospital, Department of Radiation Oncology, ایران, tarbiat modares university, Department of Statistical, ایران
 
     
   
Authors
  
 
 

Copyright 2023
Islamic World Science Citation Center
All Rights Reserved