Dosimetric characterization of a high dose rate 192I source for brachytherapy application using Monte Carlo simulation and benchmarking with thermoluminescent dosimetry
|
|
|
|
|
|
|
|
نویسنده
|
Babaei Bidmeshki N. ,Sohrabpour M. ,Mahdavi S.R.
|
|
منبع
|
international journal of radiation research - 2014 - دوره : 12 - شماره : 3 - صفحه:265 -270
|
|
چکیده
|
Background: the purpose of this project was to derive the brachytherapy dosimetric functions described by american association of physicists in medicine (aapm) tg-43 u1 based on high dose rate 192i sources. materials and methods: the method utilized included both simulation of the designed polymethyl methacrylate (pmma) phantom using the monte carlo of mcnp4c and benchmarking of the simulation with thermoluminescent (tl) dosimeters. results: the obtained results for the radial dose function and anisotropy function showed nominal errors of less than 3% between tl measurements and the mcnp4c results. conclusion: it may be concluded that due to small observed errors and the large uncertainty associated with the high dose gradients near the source point the simulation results can be used for dose estimation.
|
|
کلیدواژه
|
Brachytherapy ,192Ir flexisource ,MCNP calculation ,radiation dosimetry
|
|
آدرس
|
sharif university of technology, Department of Energy Engineering, ایران, sharif university of technology, Department of Energy Engineering, ایران, tehran university of medical sciences tums, Faculty of Medicine, Department of Medical Physics, ایران
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|